The Rigel 62353+ is the smallest and most flexible electrical safety analyzer on the market. It provides an accurate and fast solution for testing in accordance with the IEC 62353 safety standard.Offering a range of tests without the need for mains power, the 62353+ is capable of conducting battery-powered insulation and earth/ground bond testing.
This offers complete freedom to move around in often crowded rooms during the installation process, eliminating the inconvenience of trailing mains cables.The 62353+ has low voltage insulation resistance tests (50V DC and 100V DC) to complement the 250V DC and 500V DC insulation tests that were present in its predecessor, the Rigel 62353.
These tests conform to growing manufacturer’s test requirements for increasingly popular equipment running on 24V DC and 48V DC, such as operating tables, lights and mobile X-rays.

Routine testing of medical electrical equipment

Service tool for performance testing

Asset management

Fast and efficient testing of IEC leads

Bond testing on (medical) installations and non-medical equipment
Earth Continuity
Method
2 wire technique, using ‘zero’ lead function.
Test Current
>+200mA, -200mA DC into 2 ohms
Max Test Voltage
4-24V RMS o/c
Measuring Range (low range)
Measuring Range (mid range)
Measuring Range (high range)
PASS/FAIL Threshold
0.2 ohm
Accuracy
± 3% of reading + 10 m ohms
Insulation Resistance
Measurement
Voltage
50 / 100 / 250 & 500V DC @1mA
Range (low) @ 50V DC
0.01Mohms – 10 Mohms
Range (low) @ Above 50V DC
0.01Mohm – 20Mohm
Accuracy (low range)
± 5% of reading +2 counts
Range (high range) @ 250V DC
20Mohm – 50Mohm
Range (high range) @ 500V DC
20Mohm – 100Mohm
Accuracy (high range)
±10% +2 counts
Resolution
0.01Mohms
Direct Leakage Measurement
Measuring Range
Accuracy
± 5% or reading +2 counts
Mains on A.P. voltage
F-type only @ 110% of mains
Measuring Device
As per IEC 60601-1 requirements
Measurement Type
Differential Leakage Measurement
Measuring Range
Accuracy
±5% of reading + 5 counts
Measurement/display resolution
1μA
Measurement Type
True RMS
Measuring Device
Alternative Leakage Measurement
Test Voltage
Test Current
3.5 mA current limited
Measurement Range
4μA to 9999μA
Measurement Resolution
1μA
Measurement Accuracy
±5% of reading + 2 counts
Measurement Type
True RMS
Measuring Device
As per IEC 60601-1
General Specification
Mains Power
Battery
6 x 1.5V AA
Weights
1.6 kg / 3.5 lbs including batteries
Size
270 x 110 x 75 mm / 10.5 x 4 x 3”
Operating Conditions
0°C – 40°C
Storage Environment
-15’C – +60’C / 5’F – 140’F
Environmental Protection
IP40
Power Measurement
Range
Accuracy
±10% + 2 counts
IEC Mains Lead Test
Test Duration
Test
Mains Outlet Test
Input voltage range
Max current
16A
Measures
L – E, N – E
Accuracy
± 5% of reading + 2 counts
Standard Accessories (supplied with 288+)
Calibration certificate
Earth/ground bond clip lead
Patient Applied part module
10 Applied part adaptors
Detachable 2 meter mains cable
Bluetooth USB dongle
Electronic instruction manual